All Services/SVC-001
ANSI S3.6 / IEC 60645-1

Audiometer Calibration

ANSI S3.6 & IEC 60645-1 Electroacoustic Calibration

Frequency Range
125 Hz – 16 kHz
SPL Accuracy
±0.5 dB
Frequency Accuracy
±1%
SVC-001
Overview

Service Summary

Audiometer calibration is the process of verifying and adjusting an audiometer's output levels, frequency accuracy, and linearity against internationally recognized reference standards. Medology performs full electroacoustic calibration of pure-tone, speech, and diagnostic audiometers to ANSI S3.6-2018 and IEC 60645-1 specifications using NIST-traceable measurement equipment.

Frequency Range
125 Hz – 16 kHz
SPL Accuracy
±0.5 dB
Frequency Accuracy
±1%
THD Limit
< 2.5% (ANSI)
Crosstalk
≥ 70 dB attenuation
Cert. Turnaround
Same day
Importance

Why It Matters

  • OSHA, JCAHO, and state health departments require periodic calibration — typically annually — for audiometers used in clinical or occupational settings.

  • An out-of-tolerance audiometer produces inaccurate thresholds, leading to misdiagnosis, missed hearing loss, or incorrect hearing aid fittings.

  • Calibration certificates are required for accreditation inspections and must document traceability to NIST reference standards.

  • Regular calibration extends instrument lifespan by identifying transducer degradation and mechanical wear before they cause patient-facing errors.

Supported Equipment

Instruments & Platforms Covered

Diagnostic Audiometers

  • Interacoustics AC40 / AC440
  • Grason-Stadler GSI 61 / GSI 61e
  • Natus Aurical FreeFit
  • Otometrics MADSEN Astera²

Screening Audiometers

  • Welch Allyn AS-3D
  • Maico MA 41
  • Inventis Piano
  • MedRx AVANT A2D+

Portable / Handheld

  • Oscilla SM-950
  • SHOEBOX Audiometer
  • Earscan ES3
  • Amplivox 116

Don't see your model? Contact us — we service virtually all clinical audiology equipment.

Process

Service Workflow

01

Intake & Serial Verification

Equipment is received, serial numbers verified against service history, and physical condition documented before any testing begins.

Avg. 10 min
02

Electroacoustic Baseline Measurement

Output levels, frequency, linearity, and noise floor are measured on a calibrated coupler and recorded against ANSI S3.6 tolerances.

Using B&K 4153 coupler
03

Adjustment & Correction

Any parameters found outside tolerance are adjusted using manufacturer software or hardware trimmers and re-measured to confirm correction.

Critical step
04

Final Validation & Certification

Post-correction measurements confirm all parameters are in-tolerance. A NIST-traceable calibration certificate is issued digitally.

Same-day cert
Compliance

Standards & Regulations

Audiometer calibration must comply with a hierarchy of standards to satisfy regulatory, accreditation, and clinical requirements.

ANSI S3.6-2018Audiometers

Defines tolerances for pure-tone output, frequency accuracy, linearity, THD, and noise.

IEC 60645-1Pure-Tone Audiometers

International equivalent of ANSI S3.6; required for ISO 17025-accredited calibration.

ISO 17025Lab Accreditation

Framework under which Medology's calibration laboratory operates and issues traceable certificates.

OSHA 1910.95Occupational Noise

Mandates annual audiometer calibration for occupational hearing conservation programs.

Every calibration certificate includes the measurement date, technician ID, equipment serial number, reference standard used, and all measured values with pass/fail status against ANSI tolerances.

FAQ

Common Questions

Have a question not answered here? Contact our team.

Ready to proceed?

Audiometer Calibration — Request Service

Submit a request online or call for immediate assistance. Certificates issued same day.