All Services/SVC-002
IEC 60645-5 / ANSI S3.39

Tympanometer Calibration

IEC 60645-5 Middle Ear Analyzer Calibration

Probe Tone
226 Hz / 678 Hz / 1 kHz
Level Accuracy
±1.5 dB
Pressure Range
-400 to +200 daPa
SVC-002
Overview

Service Summary

Tympanometer calibration verifies probe tone frequency and level accuracy, equivalent ear canal volume measurement, and middle ear pressure sweep performance against IEC 60645-5 and ANSI S3.39 standards. Accurate tympanometry is foundational to diagnosing otitis media, Eustachian tube dysfunction, ossicular chain disorders, and sensorineural hearing loss.

Probe Tone
226 Hz / 678 Hz / 1 kHz
Level Accuracy
±1.5 dB
Pressure Range
-400 to +200 daPa
Pressure Accuracy
±10 daPa
ECV Range
0.1 – 5 cm³
Cert. Turnaround
Same day
Importance

Why It Matters

  • Miscalibrated probe tones produce erroneous compliance peaks, leading to false Type B tympanograms or missed perforation findings.

  • Pressure sweep accuracy directly affects peak pressure measurement — critical for pediatric screening protocols and tube placement decisions.

  • IEC 60645-5 calibration is required for instruments used in ISO 17025-accredited laboratory settings and JCAHO-accredited hospitals.

  • Multi-frequency tympanometry (MFT) instruments require calibration across 226 Hz, 678 Hz, and 1000 Hz probe tones independently.

Supported Equipment

Instruments & Platforms Covered

Immittance Systems

  • Interacoustics AT235 / AT235h
  • Grason-Stadler TympStar Pro
  • Otometrics TITAN / Impedance Station
  • Natus HearStation

Combination Units

  • Interacoustics Equinox 2.0
  • Otometrics AURICAL Aud
  • GSI Audera
  • MedRx AVANT Tympanometer

Portable Tympanometers

  • Welch Allyn MicroTymp 4
  • Maico MT 10
  • Inventis Timpany
  • PATH medical Sentiero

Don't see your model? Contact us — we service virtually all clinical audiology equipment.

Process

Service Workflow

01

Physical & Probe Inspection

Probe tip, port, and seal integrity checked. Probe tone output measured on a calibrated 0.5 cm³ cavity at 226 Hz, 678 Hz, and 1000 Hz.

Cavity measurement
02

Pressure System Verification

Pressure pump linearity and range verified across -400 to +200 daPa using a calibrated pressure reference and digital manometer.

±10 daPa tolerance
03

Compliance & ECV Measurement

Acoustic admittance measurement accuracy verified using calibrated cavity volumes. ECV readings compared against known-volume standards.

Critical step
04

Calibration Report Issuance

All measurements documented against IEC 60645-5 tolerances. NIST-traceable certificate issued digitally with full measurement tables.

Same-day cert
Compliance

Standards & Regulations

Tympanometer calibration requirements are defined by IEC 60645-5 and referenced by ASHA guidelines for clinical immittance audiometry.

IEC 60645-5Middle Ear Analyzers

Defines calibration requirements for probe tone level, frequency, pressure range, compliance measurement accuracy, and ECV.

ANSI S3.39Immittance Instruments

ANSI equivalent standard covering acoustic immittance instruments used in the United States.

ASHA 1997Clinical Guidelines

ASHA guidelines on audiologic screening reference IEC 60645-5 compliance for tympanometry used in pediatric protocols.

Calibration certificates document probe tone frequencies measured, level accuracy at each frequency, pressure range and linearity, ECV accuracy, and overall pass/fail status against IEC 60645-5 Table 1 tolerances.

FAQ

Common Questions

Have a question not answered here? Contact our team.

Ready to proceed?

Tympanometer Calibration — Request Service

Submit a request online or call for immediate assistance. Certificates issued same day.